SOP for Operation of Polarimeter


 


SOP for Operation of Polarimeter

1.0              Purpose

The purpose of this SOP is to lay down operational procedure of Polarimeter.

2.0              Scope

This procedure is applicable to Quality Control Department.

3.0              Responsibility

Officer QC

Asst. Manager QC

Manager QC

4.0              Accountability

Manager QA

5.0              Terms and Definitions

N/A

6.0              References

N/A

7.0              Precautions

7.1              Wear mask and gloves while operating the instrument.

7.2              Check that no air bubble is present and the solution is not turbid.

7.3              Before starting the test wait till sodium lamp warms up.

8.0              Procedure

8.1              Switch on the instrument using “Power” button.

8.1.1        Wait till sodium lamp warm and intensity of light reach its maximum for about 15   minutes.

8.1.2        Fill the tube with distilled water/blank carefully so that no bubble form and place the lens.

8.1.3        If any bubble is present transfer it to the bulb and place the cell in the chamber cover it and observe the dark and light band rotate the knob below the eye piece check for zero.

8.1.4        Take out the observation tube, discard the blank and rinse it three to four times with the sample solution.

8.1.5        Now fill the tube with sample solution carefully to prevent bubble formation and place the lens.

8.1.6        Again watch through eye piece and match the difference of two bands by rotating the knob clockwise or anticlockwise.

8.2              Note the reading where difference got minimum i.e two halves become same.

8.3              Put the value in the formula and get the results.

                                            Formula                      Degree of Rotation                x 100

                                                                   Length of tube x Conc. Of solution

8.3.1        If the knob is rotated clockwise from zero reading will be (+) degree and if it is anticlockwise from zero then it will be subtracted from 180 and a sign (-) will be     written with value. 

8.4              Record the readings in the Polarimeter log book.

8.5              Switch off and unplug the instrument and clean properly.

 

9.0              Abbreviations

9.1              SOP     Standard Operating Procedure

9.2              QC       Quality Control

9.3              QA       Quality Assurance

9.4              Asst      Assistant

10.0          Annexure

N.A