Calibration of UV/Visible Spectrophotometer

 

Calibration of UV/Visible Spectrophotometer

Calibration of UV/Visible Spectrophotometer

¨  Calibration of UV/Visible Spectrophotometer involves following parameters

¨  Absorption Cells

¨  Control of Absorbance

¨  Resolution Power

¨  Limit of Stray Light

¨  Linearity

¨  Control of Wavelength

Procedure

¨  Properly check the instrument connection and then switch on.

¨  When the instrument gets on properly then start calibration step by step for checking the above mentioned parameters one by one properly and record the data.

Absorption Cell

¨  Properly clean up the cell with methanol and remove any spot if present.

¨  Now rinse the cell with distilled water and then fill with distilled water.

¨  Measure the apparent absorbance at 240 nm for quartz cell and 650 nm for glass cell.

¨  The apparent absorbance observed should not be more than 0.093 for 1 cm quartz cell (UV regions) and 0.035 for 1 cm glass cell (Visible region).

¨  Now rotate the cell in its holder at 180◦ and measure the absorbance again and record the data.

Acceptance Criteria

¨  The difference in absorbance after rotation should not be more than 0.005 from its initial absorbance.

Control of Absorbance

¨  For the Calibration of UV region absorbance prepare 60ppm solution of Potassium dichromate solution using 0.005 M sulphuric acid as follow.

¨  Take 60 mg of potassium dichromate previously dried up to constant mass at 130◦C.

¨  Transfer it into a 1000 ml volumetric flask and dissolve in sufficient quantity of 0.005 M sulphuric acid and then make final volume with the same solvent.

¨  Measure the absorbance at 235 nm, 257 nm, 313 nm and 350 nm using 0.005 M sulphuric acid as blank.

¨  For the calibration of Visible region absorbance prepare 600ppm solution of potassium dichromate in 0.005 M sulphuric acid.

¨  Measure the absorbance of 600ppm solution of potassium dichromate in 0.005 M sulphuric acid at 430 nm using 0.005 M sulphuric acid as blank.

Acceptance criteria:

Wavelength (nm)

Specific absorbance

Limit

235

124.5

122.9 --- 126.2

257

144.5

142.8 --- 146.2

313

48.6

47.0 --- 50.3

350

107.3

105.6 --- 109

430

15.7 – 16.1

 

Resolution Power

¨  To calibrate Resolution prepare a 0.02% V/V solution of toluene in hexane.

¨  Record the spectrum of the above said 0.02% toluene solution in the range of 250 nm to 300 nm using hexane as blank.

Acceptance Criteria:

¨  Ratio of the absorbance maximum at 269 nm and minimum 265 nm should not be less than 1.5.

Limit of Stray Light

¨  For this parameter prepare a 1.2% w/v solution of potassium chloride in distilled water.

¨  Record the absorbance at 198 nm, 200nm and 202 nm using distilled water as blank.

Acceptance Criteria:

¨  Absorbance should not be less than 2.0

Linearity

¨  For linearity prepare 100 ppm, 80ppm, 60ppm, 40ppm and 20 ppm solution of previously dried potassium dichromate at 130◦c up to constant mass in 0.005 M sulphuric acid.

¨  Measure the absorbance at 257 nm using 0.005 M sulphuric acid as blank.

¨  Plot a graph between absorbance vs concentration.

Acceptance criteria:

¨  R-square should not be less than 0.999

Control of Wavelength

¨  Prepare 4.0% holmium oxide solution in 1.4 M perchloric acid by dissolving 1.0 g of holmium oxide in 25 ml of 1.4 M perchloric acid.

¨  Check the wavelength at 241.15nm, 287.15nm, 361.5nm, 486nm and 536.3 nm for the peak detection of holmium oxide using 1.4M perchloric acid as blank.

Acceptance Criteria:

¨  Tolerance limit should be ±1 for 200nm to 400nm and ±3 for 400nm to 800nm.

Out of Calibration

¨  In case of failure of any of the parameter the instrument should be labeled as “OUT OF CALIBRATION” and should be repaired / serviced.

¨  After repairing or servicing the instrument should be again calibrated.

Post a Comment

2 Comments