Calibration of Friability Test Apparatus
Calibration of Friabilator
¨ Calibration
of friability test apparatus i.e Friabilator include following parameters.
¨ Calibration
of Revolution Per Minute
¨ Calibration
of Timer
Calibration of RPM
¨ Switch
on the instrument and check that is the instrument is ready to be operate.
¨ Set
the RPM at value of 25.
¨ Press
start button and operate the instrument for 4 minutes and then stop.
¨ Count
number of revolutions manually.
¨ Calculate
number number of revolution per minute using following formula
¨ RPM=
Number of revolutions
total time (Minutes)
¨ Record
the data obtained.
¨ Repeat
the same procedure for 50RPM and 100RPM.
Calibration of Timer
¨ Switch
on the instrument and check that is the instrument is ready to be operate.
¨ If
the instrument is ready to be operate then set the timer at 4 minutes.
¨ Take
a pre-calibrated stopwatch.
¨ Start
the instrument and pre-calibrated stopwatch simultaneously.
¨ Observe
the time on the instrument and stopwatch keenly.
¨ Record
the data on calibration sheet.
¨ Repeat
the same for 8 to 12 times.
Acceptance Criteria
¨ RPM----
25±1/min
¨ Timer-----
±5 seconds
If both the test parameters were find within the acceptance range instrument should be tagged as “calibrated”. In case of any failure concerned person should be informed.