Calibration of HPLC



Calibration of High Performance Liquid Chromatography

Calibration of HPLC

       To perform calibration of HPLC following parameters should be considered.

       Leakage Test/Pressure Test

       Flow Rate Calibration

       Drift and Noise Calibration

       Column Oven Calibration

       Linearity and Reproducibility of Injection volume

       Detector Linearity response

Leakage Test/Pressure Test

       Switch on the instrument and check it if the instrument is ready for calibration proceed further for Leakage/Pressure test.

       Place the suction tube in HPLC grade water and turn on the pump.

       Allow the flow of mobile phase at the rate of 0.1ml/min for 5 minutes.

       Now block the outlet of the pump with a screw.

       Set the maximum pressure usually 300 bar.

       Record the gradual increase in pressure till it reach the maximum limit.

       The pressure rise should be smooth if it shows any fluctuation or didn’t rise up to the set limit there should be possible leakage in the system.

       Repeat the procedure for 2 more times.

       Record the data in the calibration log book.

       In case of any failure of the test inform the concern authoritative person.

Flow Rate Calibration

       When the system passes the pressure/leakage test proceed for the Flow Rate Calibration.

       Connect stainless steel union instead of column with the instrument.

       Properly degas mobile phase (Water) and kept in the reservoir.

       Purge the system at flow rate of 2ml/min for 5 minutes so that any bubble present in the lining is removed.

       Take 10ml clean and dried glass calibrated cylinder and keep at the outlet tube in such a way that drops from the outlet falls directly into cylinder without touching walls of the cylinder.

       Set the flow rate 0.5 ml/minute and start the pump and allow to run for 15 minutes.

       Now place the 10ml cylinder at the outlet tube and start pre-calibrated stopwatch simultaneously.

       Observe the volume rise in cylinder when the volume reach up to the 10ml mark stop the stopwatch and remove the cylinder.

       Record the data on calibration sheet.

       Repeat the same procedure for flow rate of 1.0 ml/minute and 2.0 ml/minute.

Acceptance Criteria

       Time observed should be ±2.0% of the theoretical time.

S.No

Flow rate

Volume delivered

Time observed

Limit

1

0.5 ml/minute

10 ml

1176—1224 seconds

2

1.0 ml/minute

10 ml

588—612 seconds

3

2.0 ml/minute

10 ml

294—306 seconds

 

Drift and Noise Calibration

Chromatographic Conditions

Detector                UV

Wavelength          254

Column                  C18 (250×4.6) 5 micron

Flow rate               1.0 ml/minute

Injection volume  0.0

Mobile Phase       Water : Methanol

                                    70   :   30

       Preparation of Mobile Phase

       Thoroughly mix 700 ml of HPLC grade water and 300 ml of Methanol HPLC grade.

       Now filter mobile phase through 0.45 micron filter paper and degas it properly with the help of sonicator.

Procedure

       Run the system as per the approved SOP.

       Purge the system at flow rate of 2 ml/minute for 5 minutes.

       Set flow rate 1.0 ml/minute and allow it to run for 15 minutes to equilibrate the system.

       Run blank injection for 15 minutes, and complete 6 replicates.

       Process the data and record the results.

Acceptance Criteria

       Noise should less than or equal to 60 micro AU

       Drift should be less than 10 milli AU/Hour

Column Oven and Sample Cooler calibration

       Calibration of column oven can be carried out at temperature of 15◦c, 30◦c and 60◦c.

       Pre-calibrated digital thermometer should be used during calibration of column oven.

       Set the temperature of the column oven at 15◦c and put the probe of calibrated thermometer inside the oven.

       Observe the temperature display for at least 15 to 20 minutes so that the temperature stabilize.

       After stable readings record the data on calibration sheet.

       Repeat the same procedure for 30◦c and 60◦c and record the data in the same way.

       Calibration of sample cooler can also be carried out in the same way.

Acceptance Criteria

       The set temperature should be within the range of ±2◦c.

Linearity and Reproducibility of Injection Volume

Chromatographic Conditions

Detector                UV

Wavelength          254

Column                  C18 (250×4.6) 5 micron

Flow rate               1.0 ml/minute

Injection volume  20.0 micron

Mobile Phase       Water : Methanol

                                    70   :   30

Preparation of Mobile Phase

       Thoroughly mix 700 ml of HPLC grade water and 300 ml of Methanol HPLC grade.

       Now filter mobile phase through 0.45 micron filter paper and degas it properly with the help of sonicator.

Preparation of standard solution

       Take 100 ml clean and dry calibrated volumetric flask and add to it 0.5ml Benzene and 0.5 ml Toluene.

       Make up volume with methanol and mix well.

Procedure

       Check the instrument that is it ready for operation.

       Flush the column at a flow rate of 1.0 ml/minute.

       Check the set parameters required for calibration of injection volume.

       Allow the mobile phase to run for 20, 30 minutes to equilibrate the column and steady baseline is obtained.

       For reproducibility of injection volume checking inject 6 replicates of the standard solution at injection volume of 20 micron.

       Process the method and record the data on the calibration table.

       Find out the RSD of the 6 replicates.

       For linearity of injection volume inject 3 replicates of injection at different injection volume i.e 10, 15, 20, 15 and 30 micron.

       Fine out the correlation coefficient “r2” for 5 levels of injection volume.

       Record the data on the calibration sheet.

Acceptance Criteria

       For reproducibility RSD should be less than 2.0%

       Ratio of the average of two areas should be less than 3.0%

       For linearity of injection volume correlation coefficient “r2” should be greater than 0.999

Injection

Area of Benzene

Area of toluene

Ratio of the two areas

1

2

3

4

5

6

Average of the ratio

Mean of areas

RSD of area

 

Injection Volume

Injection 1 Area

Injection 2 Area

Injection 2 Area

Mean

RSD%

Benzene

Toluene

Benzene

Toluene

Benzene

toluene

10 ml

15 ml

20 ml

25 ml

30ml

 

Detector Linearity Response Calibration

Chromatographic Conditions

Detector                UV

Wavelength          254

Column                  C18 (250×4.6) 5 micron

Flow rate               1.0 ml/minute

Injection volume  20.0 micron

Mobile Phase       Water : Methanol

                                    70   :   30

Preparation of Mobile Phase

       Thoroughly mix 700 ml of HPLC grade water and 300 ml of Methanol HPLC grade.

       Now filter mobile phase through 0.45 micron filter paper and degas it properly with the help of sonicator.

Preparation of standard solution

       Solution 1

       Take 100 ml clean and dry calibrated volumetric flask and add to it 0.5ml Benzene and 1.0 ml Toluene.

       Make up volume with methanol and mix well.

       Solution 2

       Take 100 ml clean and dry calibrated volumetric flask and add to it 1.0 ml Benzene and 1.0 ml Toluene.

       Make up volume with methanol and mix well.

       Solution 3

       Take 100 ml clean and dry calibrated volumetric flask and add to it 1.5ml Benzene and 1.0 ml Toluene.

       Make up volume with methanol and mix well.

Procedure

       Check the instrument that is it ready for operation.

       Flush the column at a flow rate of 1.0 ml/minute.

       Check the set parameters required for calibration of injection volume.

       Allow the mobile phase to run for 20, 30 minutes to equilibrate the column and steady baseline is obtained.

       Operate the instrument as per approved SOP

       Inject the standard solutions one by one in triplicate.

       Process the method and record the data in calibration sheet.

       Calculate the ratio of benzene and toluene area and find out the average value.

       Calculate the correlation coefficient “r2” of the average values in response to three levels

       Plot the curve of average area ratio of benzene to toluene vs  concentration.

       Find out the RSD of the retention time.

Acceptance criteria

       RSD of the retention time should not be more than 2.0%

       Correlation coefficient should NLT 0.999

Solution No.

Benzene Area and RT

Toluene Area and RT

Ratio of Benzene to Toluene Area

RT

Area

RT

Area

Solution 1

Mean

RSD

NA

NA

 

Solution No.

Benzene Area and RT

Toluene Area and RT

Ratio of Benzene to Toluene Area

RT

Area

RT

Area

Solution 2

Mean

RSD

NA

NA

 

Solution No.

Benzene Area and RT

Toluene Area and RT

Ratio of Benzene to Toluene Area

RT

Area

RT

Area

Solution 3

Mean

RSD

NA

NA

 

       If All the parameters were found within the acceptance range the Instrument should be tagged as “calibrated”

       In case of any failure the concerned person should be informed.

Post a Comment

2 Comments

  1. tumbs up for you pharmacytimess

    ReplyDelete
  2. well explained. thanks for sharing such kinda stuff

    ReplyDelete